Hemoglobin A1c (HbA1C) Turbidimetric Immunoassay kit

Hemoglobin A1c (HbA1C) Turbidimetric Immunoassay kit

  • INTRODUCTION

    HbA1c is the term known as glycated hemoglobin. Throughout the whole life cycle of red blood cell, HbA1c is formed continuously by the adduction of glucose to the N-terminal of the hemoglobin beta chain. Formation of the sugar-Hb linkage indicates the presence of excessive sugar in the bloodstream, often indicative of diabetes. Therefore, measurement of %HbA1c can also be used to determine three-month average blood sugar level since the average lifespan of a red blood cell is four months. HbA1c level in diabetic subjects will be elevated by 2-3 fold over the levels found in normal individuals. The cut-off value for %HbA1c is 6.5%. The higher the %HbA1c, the higher chance the diabetic subjects are going to develop complications.

     

    HbA1c PETIA kit developed by IMD can accurately measure %HbA1c in human whole blood sample.

  • PRINCIPLE OF THE ASSAY

    This assay is a turbidimetric immunoassay for the quantitative measurement of %HbA1c in human whole blood.  A standard or sample is added into a cuvette and mixed with the reaction buffer R1. Total hemoglobin and HbA1c have the same unspecific absorption rate to latex particles. After a short incubation, the test reagent R2 is added into the cuvette and mixed. Latex-HbA1c-mouse anti-human HbA1c antibody complex is formed. Agglutination is formed when goat anti-mouse IgG polyclonal antibody interacts with the monoclonal antibody. The extent to which the microparticles aggregate is quantified by the amount of light scattering measured as absorbance by a chemistry analyzer. The %HbA1c in unknown samples can be interpolated from a reference curve using the standards provided.

  • ASSAY PROCEDURE

    Assay procedures may vary depending on the automated chemistry analyzer to be used. A general example of assay procedures is stated as follow:

    • Dispense 150µl of R1 into a clean cuvette
    • Add 4µl of sample and incubate at 37˚C for 5 minutes
    • Further add 50µl of R2
    • Read change of absorbance at 660 nm for 5 - 8 minutes after the addition of R2
    • Calculate the %HbA1c in unknown sample by interpolation from a reference curve using the standards provided

     

    Caution: Different lot of R1, R2, standard and control should not be mixed for use

     

  • PACKING SPECIFICATIONS

    Cat. No.

    Size

    Hemolysis reagent

    Approximately tests

    51A1c-05

    R1: 15ml, R2: 5ml

    50ml

    100

    51A1c -10

    R1: 30ml, R2: 10ml

    100ml

    200

    51A1c -20

    R1: 60ml, R2: 20ml

    200ml

    400

    51A1c -50

    R1: 150ml, R2: 50ml

    500ml

    1000

    51A1c -100

    R1: 300ml, R2: 100ml

    1000ml

    2000

  • OTHER MATERIALS REQUIRED

     

    Product Catalog Number Format Method
    Calibrator 51Ac1-S1 4 x 0.5 ml  
    Control 51Ac1-C1 2 x 0.5 ml  

     

價格自 $275.00

如需批量訂單(數量:5 個或更多),請聯繫我們獲取報價